TS70 Series Spectrophotometer Vertical Multi-Function Accessory Installation & Operation Instructions
When measuring with the vertical multi-function accessory, the instrument is used in an inverted position. You need to rotate the instrument software interface upside down, or connect the instrument to a computer via a USB cable and perform measurements using the corresponding PC software.
Installation steps are shown in the diagrams below:
As shown in Figure 1: Stand the accessory upright, use a hex wrench to remove the two screws, and take off the movable block.

As shown in Figure 2: Install the flat-bottomed measuring aperture (pointed-bottom aperture is not suitable) onto the instrument. Complete the necessary calibration and settings, then slide the instrument into the accessory in the direction of the arrow (make sure the instrument power switch faces outward).

As shown in Figure 3: Slide the movable block back into the accessory in the direction of the arrow and tighten the two screws. A large head screw may also be inserted into the position shown in the diagram for additional fixation.

As shown in Figure 4: Carefully place a petri dish containing the sample at the center of the measuring port, cover with the light shield, and measurement can begin. Solid samples may also be placed directly on the measuring port for testing.

Notes: Since the measuring port faces upward, please handle with extreme care during measurement. Do NOT allow any objects (liquids, dust, small particles) to fall into the measuring port! In particular, liquid ingress may cause permanent damage to the instrument. Remove the sample promptly after measurement and replace the light shield to prevent dust contamination.